Drive STINGRAY Operator's Manual Page 39

  • Download
  • Add to my manuals
  • Print
  • Page
    / 58
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 38
Host Interface
5 - 1
SECTION FIVE: HOST INTERFACE
In this section:
Host command quick reference (this page)
Message considerations (p. 5-2)
Commands and queries (p. 5-5)
SCPI error codes (p. 5-10)
When a command is sent to the Coherent StingRay laser, the
parameter for the command is stored in internal persistent
memory, which has a logic cell life of 10 thousand cycles. The cell
life sets the limits for repetitive commands sent to the Coherent
StingRay laser.
This only applies to commands and not queries.
Host Command
Quick
Reference
The following table gives a brief description of all host commands
and queries. For detailed information about a specific command or
query, go to the page referenced in the right-hand column.
Table 5-1. Host Command Quick Reference (Sheet 1 of 2)
Command Description
Page
No.
*IDN? Device ID query 5-5
SYST:CDRH Enables/disables 5-second CDRH delay 5-5
SYST:CDRH? Queries CDRH delay state 5-5
SYST:COMM:BAUD Sets serial communication baud rate 5-6
SYST:COMM:BAUD? Queries serial communication baud rate 5-6
SYST:COMM:HAND Enables/disables SCPI handshaking 5-6
SYST:COMM:HAND? Queries SCPI handshaking state 5-6
SYST:COMM:PROM Enables/disables interactive prompt 5-6
SYST:COMM:PROM? Queries interactive prompt state 5-6
SYST:DIOD:HOUR? Queries laser diode usage hours 5-6
SYST:FAUL? Queries system fault 5-6
SYST:INF:CDAT? Queries factory calibration date 5-7
Page view 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 57 58

Comments to this Manuals

No comments