Environment
Coherent StingRay lasers are suitable for regular indoor and outdoor
use and function normally when the following environmental condi-
tions are met:
Altitudes up to 2000 m.
Environments where the maximum relative humidity (RH) is
80% (for temperatures up to 31°C). Note that above 31°C, the
RH decreases linearly from 80 to 50% (at 50°C).
Environments in which the diode is soaked –10 to 50°C. In
warmer environments, a heatsink or a thermoelectric cooler
should be used to minimize the heat build-up. In extremely
cold environments, care should be taken to maintain the laser
above –10°C at all times.
NOTICE!
As with all semiconductor materials, avoid prolonged or
repeated exposure to electrostatic charges or water droplets. All
Coherent StingRay lasers are designed with ESD protection.
Operate the laser in an environment in which there is normal aera-
tion.
Table 2-4. Modulation Characteristics and States
Modulation F
max
Direct Indirect
Analog 500 KHz
0 to 0.5 VDC 4.5 to 5 VDC
OFF
0.5 to 5 VDC 0 to 4.5 VDC
LINEAR REGION
TTL 100 KHz
0 to 1 VDC 4 to 5 VDC
OFF
4 to 5 VDC 0 to 1 VDC
ON
1 to 4 VDC 1 to 4 VDC
UNDEFINED
Fast TTL 2 MHz
0 to 1 VDC 4 to 5 VDC
OFF
4 to 5 VDC 0 to 1 VDC
ON
1 to 4 VDC 1 to 4 VDC
UNDEFINED